Client Background
Nuvai Medical Technologies needed its medication-verification demo to do more than read a vial. The same offline device had to inspect the drawn syringe and check its volume against the preparation order.
This case study covers the syringe-measurement workstream within the same first-phase Nuvai engagement, not a separate clinical deployment.
The Problem
Recognizing a syringe is not the same as measuring what it contains. The system needed to locate the plunger, interpret the printed scale and know when the image could not support a dependable measurement.
Different syringe sizes have different scales, so the device needed to check the size before measuring.
Blur, glare, rotation and hand-held positioning could make an apparently plausible number unreliable.
Small draws and uncertain measurements needed an explicit review path rather than false precision.
Our Solution
NextGen trained an object detector on captured images to locate the syringe, barrel, plunger stopper and barrel tip.
The measurement pipeline straightens the barrel, checks its shape against the ordered syringe size and identifies the stopper's leading edge.
Image processing reads the printed graduations in each frame to establish its scale, then converts the plunger position to volume.
Confidence combines detection, edge sharpness and scale quality. An uncertain scale or blurred edge produces review rather than an invented number.
The operator flow gives add-or-reduce guidance for an out-of-range draw, with bounded rechecks and a review outcome when correction does not resolve it.
Results
Live syringe checks on the client's staging Jetson had a reported median time of 2.2 seconds and a maximum of 2.6 seconds in the documented sessions.
The fixed-tray approach was the stronger demo setup; hand-held edge cases remained documented limitations, not claimed solved problems.
Syringe-size disagreement, unreadable scales and draws below the measurable range have explicit review outcomes.
These are prototype engineering observations from the supplied project report, not clinical validation, regulatory clearance or a claim of patient-use readiness.